Application of a CdTe Detector for Measurements of Mammographic X-ray Spectra

Authors

  • J. C. Santos Instituto de Física, Universidade de São Paulo, SP, Brazil
  • A. Tomal Instituto de Física Gleb Wataghin, Universidade Estadual de Campinas, SP, Brazil
  • P. R. Costa Instituto de Física, Universidade de São Paulo, SP, Brazil

DOI:

https://doi.org/10.15415/jnp.2017.51009

Keywords:

CdTe detector, Mammography, x-ray spectra measurements

Abstract

This work aims to characterize mammographic x-ray beams incident and transmitted by breast phantoms (from 0 to 45 mm) composed from known proportion of glandular and adipose tissue-equivalent materials. This study was performed for mammographic x-ray beams generated by a mammography equipment using different target/filter combinations (Mo/Mo, Mo/Rh and W/Rh). It was studied the modification of spectra shape of the beams transmitted through different thicknesses of these materials. It was also evaluated the penetrability of these transmitted beams by its correlations to the HVL, which were experimentally estimated and derived from the x-ray spectra measured using a spectrometry system with a CdTe detector. The x-ray spectra transmitted by the phantom with higher density presented lower intensity than those transmitted by those with lower density, as expected. The differences between the HVL values derived from the spectra and those estimated using air kerma measurements are lesser than 6% for about 88% of the spectra measured in this work. The expected spectra variations with phantom thickness, revealed by the measured transmitted x-ray spectra, were also confirmed by HVL measurements and agree with the estimated attenuation curves.The motivation of the study was related to the robustness of the spectra as a descriptor of radiation beams and the possibility of using these transmitted spectra for dose assessment related to mammographic procedures. We can conclude that developed method is able to characterize mammographic x-ray beams making it possible the use of this kind of data for dose assessment in mammography.

Downloads

Download data is not yet available.

References

Archer, B.R., Thornby, J.I., Bushong, S.C., 1983. Diagnostic X-ray shielding design based on an empirical model of photon attenuation. Health Phys 44, 507–517.

Bernhardt, P., Mertelmeier, T., Hoheisel, M., 2006. X-ray spectrum optimization of full-field digital mammography: Simulation and phantom study. Medical Physics 33, 4337–4349.

Blough, M.M., Waggener, R.G., Payne, W.H., Terry, J.A., 1998. Calculated mammographic spectra confirmed with attenuation curves for molybdenum, rhodium, and tungsten targets. Med. Phys. 25, 1605–1612.

Boone, J.M., Fewell, T.R., Jennings, R.J., 1997. Molybdenum, rhodium, and tungsten anode spectral models using interpolating polynomials with application to mammography. Med. Phys. 24, 1863–1874.

Byng, J.W., Mainprize, J.G., Yaffe, M.J., 1998. X-ray characterization of breast phantom materials. Physics in Medicine and Biology 43, 1367.

Di Castro, E., Pani, R., Pellegrini, R., Bacci, C., 1984. The use of cadmium telluride detectors for the qualitative analysis of diagnostic x-ray spectra. Phys Med Biol 29, 1117–1131.

Geraldelli, W., Tomal, A., Poletti, M.E., 2013. Characterization of TissueEquivalent Materials Through Measurements of the Linear Attenuation Coefficient and Scattering Profiles Obtained With Polyenergetic Beams. IEEE Transactions on Nuclear Science 60, 566–571.

Hammerstein, R.G., Miller, D.W., White, D.R., Ellen Masterson, M., Woodard, H.Q., Laughlin, J.S., 1979. Absorbed Radiation Dose in Mammography. Radiology 130, 485–491.

Hernandez, A.M., Boone, J.M., 2014. Tungsten anode spectral model using interpolating cubic splines: Unfiltered x-ray spectra from 20 kV to 640 kV. Medical Physics 41.

Johns, P.C., Yaffe, M.J., 1987. X-ray characterisation of normal and neoplastic breast tissues. Physics in Medicine and Biology 32, 675.

Künzel, R., Herdade, S.B., Terini, R.A., Costa, P.R., 2004. X-ray spectroscopy in mammography with a silicon PIN photodiode with application to the measurement of tube voltage. Medical Physics 31, 2996–3003.

Künzel, R., Levenhagen, R.S., Herdade, S.B., Terini, R.A., Costa, P.R., 2008. X-ray spectroscopy applied to radiation shielding calculation in mammography. Medical Physics 35, 3539–3545.

Meyer, P., Buffard, E., Mertz, L., Kennel, C., Constantinesco, A., Siffert, P., 2004. Evaluation of the use of six diagnostic X-ray spectra computer codes. Brit. J. Radiol. 77, 224–230.

Santos, J.C., Mariano, L., Tomal, A., Costa, P.R., 2016. Evaluation of conversion coefficients relating air-kerma to H*(10) using primary and transmitted x-ray spectra in the diagnostic radiology energy range. Journal of Radiological Protection 36, 117–132.

Santos, J.C., Tomal, A., Furquim, T.A., Fausto, A.M.F., Nogueira, M.S., Costa, P.R., 2017. Technical Note: Direct measurement of clinical mammographic x-ray spectra using a CdTe spectrometer. Medical Physics, n/a-n/a.

Stumbo, S., Bottigli, U., Golosio, B., Oliva, P., Tangaro, S., 2004. Direct analysis of molybdenum target generated x-ray spectra with a portable device. Med. Phys. 31, 2763–2770.

Tomal, A., Cunha, D.M., Poletti, M.E., 2013. Optimal X-ray spectra selection in digital mammography: A semi-analytical study. IEEE Transactions on Nuclear Science 60, 728–734.

Tomal, A., Cunha, D.M., Poletti, M.E., 2014. Comparison of beam quality parameters computed from mammographic x-ray spectra measured with different high-resolution semiconductor detectors. Radiation Physics and Chemistry 95, 217–220.

Tomal, A., Santos, J.C., Costa, P.R., Lopez Gonzales, A.H., Poletti, M.E., 2015. Monte Carlo simulation of the response functions of CdTe detectors to be applied in x-ray spectroscopy. Appl Radiat Isot 100, 32–37.

Tucker, D.M., Barnes, G.T., Wu, X., 1991. Molybdenum target x-ray spectra: A semiempirical model. Med. Phys. 18, 402–407.

Downloads

Published

2017-08-07

How to Cite

(1)
J. C. Santos; A. Tomal; P. R. Costa. Application of a CdTe Detector for Measurements of Mammographic X-Ray Spectra. J. Nucl. Phy. Mat. Sci. Rad. A. 2017, 5, 91-102.

Issue

Section

Articles